11593492 enUV-Vis DetectorC�640Operation Manual
2 Safety10 C-640 Operation Manual, Version A NoticeRisk of instrument damage by wrong mains supply.• External mains supply must meet the voltage giv
3 Technical data11 C-640 Operation Manual, Version A3 Technical dataThis chapter introduces the reader to the instrument and its specifications. It c
3 Technical data12 C-640 Operation Manual, Version A3.1.2 Standard equipmentEach part listed with an order number can be ordered separately as spare
3 Technical data13 C-640 Operation Manual, Version AFEP tube inner diameter 1/16”, outer diameter 1/8”044354FEP tube inner diameter 1/32”, outer diam
3 Technical data14 C-640 Operation Manual, Version A3.4 Technical specificationsTable 3-5: Technical specificationsAverage power consumption max. 90 W
3 Technical data15 C-640 Operation Manual, Version A3.5 Configuration of the analog cinch outputa Analog Output (HI) Absorbance signal output. Volt
3 Technical data16 C-640 Operation Manual, Version A3.7 Signals and parameters3.7.1 Light intensityThe light intensity of a newly installed lamp us
3 Technical data17 C-640 Operation Manual, Version A3.7.8 ScanningScanning over the whole detection range causes a high workload on the detector and
4 Description of function18 C-640 Operation Manual, Version A4 Description of functionThis chapter explains the basic working principle of the UV-VI
4 Description of function19 C-640 Operation Manual, Version A4.1.2 Icons on the graphical displayIcon MeaningIf this icon of a small key is visible
4 Description of function20 C-640 Operation Manual, Version A4.3 The cell compartmentThe cell compartment is located on the right side of the instru
4 Description of function21 C-640 Operation Manual, Version A4.3.1 Removing the cell For a safe removal of the cell from the cell compartment procee
4 Description of function22 C-640 Operation Manual, Version A4.3.2 Mounting the cell - Enter the cell to the cell compartment, slide it onto the gr
5 Putting into operation23 C-640 Operation Manual, Version A5 Putting into operationThis chapter describes how the instrument has to be installed. I
5 Putting into operation24 C-640 Operation Manual, Version A5.2 Installing the Detector C-6405.2.1 Electrical connectionsVerify that the electrical
5 Putting into operation25 C-640 Operation Manual, Version A5.2.3 Connecting the tubingTubes with two different diameters are contained in the stand
5 Putting into operation26 C-640 Operation Manual, Version A 21- Remove the white blanking plugs from the cell and connect the column outlet to the
5 Putting into operation27 C-640 Operation Manual, Version A5.4 Controlling the detectorThe detector can be controlled in two ways:• in local contr
5 Putting into operation28 C-640 Operation Manual, Version Aparameters of each screen (e.g. through the different wavelengths). The value of a parame
5 Putting into operation29 C-640 Operation Manual, Version A5.5.2 Time constantWithin this menu the time constant in seconds can be set for all avai
Table of contents3 C-640 Operation Manual, Version ATable of contents1 About this manual . . . . . . . . . . . . . . . . . . . .
5 Putting into operation30 C-640 Operation Manual, Version A5.6 Getting information and scanning5.6.1 General informationThe General info menu prov
5 Putting into operation31 C-640 Operation Manual, Version A5.6.4 ErrorsIf any kind of error occurs during operation of the detector an error descri
6 Maintenance and repairs32 C-640 Operation Manual, Version A6 Maintenance and repairsThis chapter gives instructions on maintenance work to be perf
6 Maintenance and repairs33 C-640 Operation Manual, Version A6.3.1 Cleaning the cell windowA contaminated cell will result in a lowered light transm
6 Maintenance and repairs34 C-640 Operation Manual, Version ANoteTo change the thickness of the liquid layer (the optical path) between the two cell
6 Maintenance and repairs35 C-640 Operation Manual, Version A6.4 Replacement of the deuterium lamp! WarningDeath or serious injuries by contact with
6 Maintenance and repairs36 C-640 Operation Manual, Version A1- Loosen the screw a on the lamp holder just so far that the lamp holder becomes turna
7 Troubleshooting37 C-640 Operation Manual, Version A7 TroubleshootingProblem Caused by / Conditions SolutionAfter switching the unit on no LED on t
7 Troubleshooting38 C-640 Operation Manual, Version A7.2 Hardware error codesAll errors related to a faulty hardware part of the detector are consid
8 Shutdown, storage, transport and disposal39 C-640 Operation Manual, Version A8 Shutdown, storage, transport and disposalThis chapter instructs how
Table of contents4 C-640 Operation Manual, Version A4.1.2 Icons on the graphical display . . . . . . . . . . . . . . . . . . . .
9 Declarations and requirements40 C-640 Operation Manual, Version A9 Declarations and requirements9.1 FCC requirements (for USA and Canada)English:
8 Shutdown, storage, transport and disposal41 C-640 Operation Manual, Version A
9 Declarations and requirements42 C-640 Operation Manual, Version A
BÜCHI Labortechnik AGCH-9230 Flawil 1 / SwitzerlandT +41 71 394 63 63 F +41 71 394 65 65 www.buchi.com Quality in your hands
1 About this manual5 C-640 Operation Manual, Version A1 About this manualThis manual describes the UV-VIS Detector C-640. It provides all informatio
2 Safety6 C-640 Operation Manual, Version A2 SafetyThis chapter points out the safety concept of the instrument and contains general rules of behavi
2 Safety7 C-640 Operation Manual, Version A2.4 Safety warnings and safety signals used in this manualDANGER, WARNING, CAUTION and NOTICE are standar
2 Safety8 C-640 Operation Manual, Version AExplosive gases, explosive environmentHarmful to live-formsDevice damageWear laboratory coatWear protectiv
2 Safety9 C-640 Operation Manual, Version A2.5 Product safetySafety warnings in this manual (as described in section 2.4) serve to make the user ale