6 Operation
38 KjelFlex K-360 Operation Manual, Version C
Table 6-1: Titrator settings
Titrator present Configuration Signal ready Signal active Signal end Baud Data bit Parity Stop bit
1 e.g. Mettler Toledo DL 15/22 Low No RS232 9600 8 no 2
2 e.g. Mettler Toledo T50/Txx No No RS232 4800 8 even 1
3 e.g. Metrohm Food Titrino
DMP785 / 719 / 702
No No RS232 9600 8 no 1
4 e.g. Schott TitroLine Easy No RS232 RS232 4800 7 no 1
5 e.g. Radiometer TitraLab 840 Low High Low 9600 8 no 1
6 e.g. Metrohm 848 Plus Low High RS232 19200 8 no 1
Custom other titrators
When you have selected “Custom” you can enter the corresponding settings via the arrow buttons
and press Save to store them.
At the moment there are the following titrator types that enable a signal transfer and thus an automatic
result display in the K-360 software:
Mettler Toledo DL 15 / 22 / Txx•
Schott TitroLine Easy•
Metrohm Titrino DMP 785 / 719 / 702 / 848 Plus•
To assure a correct signal transfer, the respective titrator has to be configurated according to the
Büchi specifications.
6.4.5 Boric acid / end point titration
Receiving vessel: see chapter 5.6
Titrant: see chapter 5.6
It is recommended to use boric acid solution 4%, adjusted to pH 4.65 and standardized sulfuric acid
solution.
Titrate the distillate with sulfuric acid standard solution to the end point of pH 4.65.
Defining settings to work with a titrator
Configuration > Settings > Rec Vessel present: Yes
Configuration > Titrator: Titrator present: (1, 2, 3, 4, 5, 6, Custom)
Methods > Select the corresponding method from the method list and press Enter
> Aspiration Sample: Yes
> Aspiration Rec. Sol’n: Yes
Defining settings without a titrator
Configuration > Settings > Rec Vessel Present: No
Configuration > Titrator: Titrator present: No
Methods > Select the corresponding method from the method list and press Enter
> Aspiration Sample: Yes