4|Descriptionoffunction BÜCHILabortechnikAG
24/110 PrepChromC700
4.3.3 Controlunit
Thesystemiscontrolledbythecentralprocessingunitviaa10.1"touchscreen,
thekeyboardandmousecanbeusedoptionally.Theoperatorusesthecontrol
softwaretorunthedevice.
Thecontrolunitcentralizestheconnectionfordatatotheothermodules.
4USBconnectorsareavailable,1intherearand3ontherightsideofthe
touchscreen(cablemaximumlength=2m).
4.3.4 Pump
Fig.7:Pumpmodulecomponents
1 Checkvalve 2 Pumphead
3 Fromthe4wayaspirationvalve 4 Leaksensor
5 Chamber,pressuresensor 6 Tothemixingchamber
The2headpumpdeliversmobilephaseatflowratesbetween5–250mL/min
withlowresidualpulsation.Itisahighperformancedoublepistonpump(recip
rocalmovement).Asystem(pulleys+belt),asteppermotordrives2camson
whichleanthe2pistons.Theirmotiongeneratesacycleandthencreatesthe
aspiration.
Astaticmixingchamberattheoutletensuresasuitablemixingofsolvents
whatevertheworkingflowrateis.
Asensordetectsleakages.Thepumpisinterruptedifaleakageoccursandthe
userisinformedviathetouchscreen.
Apressuresensormonitorsthecurrentpressure.Ifthepressureishigherthan
100barthepumpwillbeswitchedofftopreventanydamagetothesystem.
NOTE
Whenworkingwithaqueoussolutions,residuesmightcollectbehindthepiston
seals.Theseresiduesmustberemovedsincetheycancontributetoafaster
wearingoftheseals.Rinsingofthepistoncleaningdiscisdescribedin7.2.4
Piston cleaning disc, cleaning [} page90]