3 Technical data
12 F-100 / F-105 / F-108 / F-114 Operation Manual, Version E
Technical data (115 V)
F-100 F-105 F-108 F-114
Power consumption (max.) 850 W 850 W 1350 W 1650 W
Supply voltage 115 VAC ±10 % 115 VAC ±10 % 115 VAC ±10 % 115 VAC ±10 %
Fuse 10 AT 10 AT 12 AT 15 AT
Frequency 60 Hz 60 Hz 60 Hz 60 Hz
Environmental conditions For indoor use only
- Temperature 5-35 °C 5-35 °C 5-40 °C 5-40 °C
- Altitude up to 2000 m
- Humidity Maximum relative humidity 80 % for
temperatures up to 31 °C decreasing
linearly to 50 % relative humidity at 40 C°
Display No display digital, resolution
0.1 °C
digital, resolution
0.1 °C
digital, resolution
0.1 °C
Overvoltage category II II II II
Degree of protection IP20 IP20 IP20 IP20
Pollution degree 2 2 2 2
Instrument dimensions
W×H×D [mm]
280×590×420 280×590×420 400×580×500 400×660×500
Weight 37 kg 40 kg 40 kg 42 kg
Cooling power at 15 °C 400 W 500 W 800 W 1400 W
Refrigerant R134 R134 R134 R134
Temperature range fix +10 °C -5 °C ... +25 °C -10 °C ... +25 °C -10 °C ... +25 °C
Accuracy ±2 °C ±1 °C ±1 °C ±1 °C
Tank volume 3 L 3 L 4.5 L 6.6 L
Hose connection 8.0 mm 8.0 mm 9.5 mm 13.5 mm
Pump flow rate 2.5 L/min 2.5 L/min 3.0 L/min 11.0 L/min
Pump pressure 0.6 bar 0.6 bar 0.6 bar 1.0 bar