6 Operation
105
K-375/376/377 Operation Manual, Version B
Aspiration Parameters
After a determination has been completed, the waste in the sample tube and/or in the receiving
vessel can automatically be aspirated and transferred into a corresponding waste container:
Click on “Aspiration sample tube“ and select Yes
to switch on the automatic aspiration of the sample
tube.
Click on “Aspiration receiving vessel” and select
Yes to switch on the automatic aspiration of the
receiving vessel.
NOTE
If you select the automatic aspiration of the sample
tube and/or the receiving vessel and the waste is
collected, we recommend the use of level detectors
for the corresponding waste container(s). The level
sensors can be configured within the Settings
Peripherals screen (See chapter “6.9.1 Settings“)
NOTE
Samples containing bigger particles should not be
aspirated as they may cause problems with the
valves (leaks and blockages).