8 Troubleshooting
144
K-375/376/377 Operation Manual, Version B
Boiling retardation/bumping/foaming
Missing digestion rods or boiling stones are used
Missing Antifoam tablet or other anti-foaming agent
Typical mistakes distillation
Samples do not get dark blue/brown after addition of NaOH
Empty NaOH tank
Air in NaOH hose
No catalyst used for digestion (only H
2O2)
Splashing during distillation or addition of chemicals
Choosen wrong sample tubes
Volume in sample tubes too high
Not enough water used for dilution
Other problems that may occur
Problem Cause Corrective measure
Nitrogen content too high Air in titration system, buret, tubes
Carry over during distillation
Wrong titrant
Error in calculation
Defective pH electrode
Dirty glassware
Refill buret
Use less volume, or increase water
volume for dilution
Use right concentration
Check calculation and concentration
of titration, molar reaction factor,
factor of titrant
Calibrate electrode, replace if neces-
sary
Only use clean glassware