Rotavapor R-250
4.8 Installation of the Descending Glass Assembly
All glassware used must be intact, with no signs of cracks,
spreading impact marks, or other damage. Inspect the glass-
ware visually before installing it.
s &IXTHEDESCENDINGDISTRIBUTORHEAD in position on the
gear head using the DN 70 EasyClamp connection.
s &IT THE COOLING WATER HOSE NIPPLES to the two or, if
appropriate, three coolers .
s 0UTTHEEXPANSIONVESSEL onto the distribution head
and secure it with the DN 40 EasyClamp connection.
s &ITTHETHREEWAYlTTING onto the receiver using the DN
25 EasyClamp connection.
s 3ECURETHECOOLERHOLDER to the two or three coolers
, as appropriate, and insert in the pivoting clamps .
s &ITTHEVACUUMCONNECTOR or the third cooler.
s &ITTHEBRANCHINGPIECE on the cooler and the ex-
pansion vessel and connect it using the EasyClamp
connection. Align the glass components in a vertical
position and secure them using the pivoting clamps .
s &ITTHEINLETTAP into the distribution head and tighten
the SVL 30 nut.
s 3CREWTHETEMPERATURESENSOR into the branching piece
.
s 4IGHTENALL%ASY#LAMPCONNECTIONSEVENLYANDINPARALLEL
s 4URNTHEPOSITIONINGNUTTORAISETHEmASKSUPPORT#ONNECT
the bracket of the receiving flask to the condensation
flask.