22
Rotavapor R-250
5 Operation
Fig. 5.1: Operating unit of the R-250
5 Operation
Make sure that the unit has been commissioned properly as
described in Chapter 4.
5.1 Layout of Operating and Display Elements
햲 Main breaker switch
햳 Lift for the bath
햴 Rotation ON/OFF
햵 Rotating speed adjustment
햶 Rotating speed display
햷 Heater ON/OFF
햸 Input: Set-point temperature for bath
햹 Temperature UP/DOWN
햺 Display for bath temperature
햻 Display for vapour temperature
햽 Aeration
When the unit is switched off or in the event of a power failure,
the bath can be lowered automatically so that the evaporating
flask is always separated from the source of heat.
5.2 Setting the Maximum Setpoint Temperatu-
re and Lowering the Bath Automatically
Setting the Maximum Setpoint Temperature
s 0RESSTHEBUTTONh3%44%-0v햸, simultaneously swit-
ching on the main breaker switch 햲. This activates the
input for setting the maximum set-point temperature.
s 4HEVALUECANBESETATANYLEVELDESIREDBYPRESSINGTHE
h$/7.vANDh50vBUTTONS햹.
s 4HEINPUTISSTOREDANDTHEUNITSWITCHEDINTOTHEOPERATING
MODEONLYAFTERTHEh!%2!4%vBUTTON햽 is then pressed.
햽
햷
햹
햸
햺
햻
햳
햵
햴
햶
햲
R-250